Development of a rapid and fully automated factor VIII inhibitor assay, insensitive to emicizumab, and a lowest level of quantification of 0.2 BU/mL

Development of a rapid and fully automated factor VIII inhibitor assay, insensitive to emicizumab, and a lowest level of quantification of 0.2 BU/mL

Bert Verbruggen 1), Nikolaus B. Binder 2), Petra J.C. van Velp 3), Robert Polenewen 1), Paul Knöbl 4), Frédéric Sobas 5), Gary W. Moore 6), 7)

1) Hemostad, Nijmegen, The Netherlands
2) Research and Development, Technoclone Herstellung von Diagnostika und Arzneimitteln GmbH, Vienna, Austria
3) Rock Consultancy, Bemmel, The Netherlands
4) Department of Medicine 1, Division of Hematology and Hemostasis, Medical University of Vienna, Vienna, Austria
5) Haemostasis Department, Hospices Civils de Lyon, Lyon, France
6) Department of Haematology, Specialist Haemostasis Unit, Cambridge University Hospitals National Health Service Foundation Trust, Cambridge, United Kingdom
7) Department of Natural Sciences, Middlesex University, London, United Kingdom

Abstract

Background

Factor (F)VIII inhibitors are measured using labor- and resource-expensive Nijmegen or Bethesda assays, which lack sensitivity for low-titer inhibitors and show high variations in quality surveys, mainly because of manual assay procedures.

Objectives

The goal of this study was the development of a fast and fully automated FVIII inhibitor assay by using recombinant (r)FVIII as substrate and dedicated equipment for execution of the test.

Methods

A new rapid, fully automated, FVIII inhibitor assay is presented, the core of which is use of full-length recombinant FVIII (rFVIII; Kovaltry, Bayer) as inhibitor substrate instead of plasma FVIII, resulting in rapid binding of inhibitors to rFVIII due to absence of von Willebrand factor. Dramatic shortening of incubation time facilitated full automation on an analyzer capable of 3 subsequent sample dilution steps and 3 reagent additions. Equal volume mixtures of sample and rFVIII (1.0 U/mL) were incubated for 10 minutes at 37 °C, whereafter remaining FVIII activity was analyzed with a kinetic chromogenic assay, allowing inhibitor activity calculation without preceding FVIII activity calibration, using a Ceveron s100 analyzer (Technoclone).

Results

Mean titer in 60 nonhemophiliacs was 0.0 BU/mL (SD, 0.1), yielding a limit of blank of 0.1 BU/mL and lower limit of quantification of 0.2 BU/mL. Analyses were performed with the new method and a Nijmegen assay in 28 inhibitor-positive clinical samples, 14 containing emicizumab and 14 without. Correlation coefficient in emicizumab-free type I inhibitor samples was 1.0. Emicizumab dependency of the method was excluded in spiking experiments with inhibitor-positive samples. Reproducibility was tested by analyzing 7 samples in 3 laboratories for 5 days, twice daily; coefficients of variation of all samples were <15%.

Conclusion

We present development data of a sensitive and specific rapid, automated FVIII inhibitor assay generating results within 20 minutes that is less resource-intensive than standard assays with potential to improve assay variability.